Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
1. Insert the pump rotor assembly into the fixture.
Figure 20,
Figure 21,
Note: Ensure the pump rotor and the fixture are engaged by introducing the
fixture’s stud into the slit of the rotor. By design, one of the rotor rollers will be