156
Troubleshooting
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
11
SPM FAILED 50%
ANALOG TEST
SPM channel of A/D
converter failed 50%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(
MODIFY PROGRAM
) if parameters
were different from the default values.
11
SPM FAILED 80%
ANALOG TEST
SPM channel of A/D
converter failed 80%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(
MODIFY PROGRAM
) if parameters
were different from the default values.
Table 9, CPU error messages
Cod
e
Text Message
Description
Suggested Resolution