Troubleshooting
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
509
SOLUTIONS AIR
DETECTOR FAULT
SLAD does not sense air
during system start-up air
detector test.
1. Use air detector test in Diagnostics to
verify problem.
2. Refer to system schematic and
check, repair, or replace all electrical
connections associated with the SLAD.
3. Replace SLAD air detector.
4. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
510
BLOOD AIR
DETECTOR FAULT
BLAD does not sense air
during system start-up air
detector test.
1. Use air detector test in Diagnostics to
verify problem.
2. Refer to system schematic and
check, repair or replace all electrical
connections associated with the BLAD.
3. Replace BLAD air detector.
4. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
Table 9, CPU error messages