Principles of Operation
55
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Note: These parameters are market dependent and are only changed if the
“Machine Market” configuration is changed.
The following flow chart detail the User interface within the protocol
configuration of the ACP215. These are market dependent and will be
available once the market configuration has been set under the machine
configuration
menu.
Note: See configuration setup procedure in
.
Table 4, Configuration parameter default settings
International market
North America market
Config parameter
Default value
Config parameter
Default value
*Disposable Checklist
Minimum Set
*Disposable Checklist
Full Set
*Modifiable Parameters
Full Set
*Modifiable Parameters
Minimum Set
Available Protocols
Frozen Blood
Available Protocols
Frozen Blood
Cell Wash Additive Option
Enabled
Cell Wash Additive Option
Disabled
Bowl Size Option
Small Bowl
Only (275ml)
Bowl Size Option
Small Bowl
Only (275ml)
*Sterile Check in Cell Wash
Enabled
*Sterile Check in Cell Wash
Disabled