Preventive Maintenance and Installation
93
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Valves Test
Valve Occlusion Test
Note: Power on the device in the Utilities mode of operation by pressing and
holding
SHAKER
while switching the power on. Hold
SHAKER
until the
UTILITIES menu appears on the display. On the UTILITIES menu, press
MODIFY PROGRAM
until the arrow points to DIAGNOSTICS, then press
START
.
1. Select the valves test by using
MODIFY PROGRAM
and
START
.
2. Open all six valves by pressing:
HIDDEN KEY#2
SHAKER
HIDDEN KEY#3
PUMP SPEED UP
PUMP START/STOP
PUMP SPEED DOWN
for the locations of hidden keys.
Test valve 1 (red)
1. Install a length 0.170” tubing through the open pinch valve 1(red) and
attach one end to a sphygmomanometer or digital pressure meter.
2. Verify that valve 1 opens and its LED is lit.
3. Press
PUMP UP ARROW
and the valve closes.
4. Verify that valve 1 closes and its LED is not lit.
5. Pressurize the sphygmomanometer, or digital pressure meter, to
200 mmHg and verify that there is not a leak as evidenced by a decrease
in pressure. Press
PUMP UP ARROW
and remove the tubing.