92
Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
seconds.
Figure 27,
Note: When properly measured, the indenter will record its peak reading when
the foot of the tool is tangent to the roller and the foot is centered on the roller
at the same time.
Figure 28,
4. Take at least 3 measurements per roller and compute the average
reading. Replace rotor assembly or the roller if either of the average
measurements fall outside of the specification for durometer or if the
roller is sticky. The specification is 63 to 75 Shore A Durometer.