130
Disassembly
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
Rear Panel Assembly
Rear Panel
Note: After removing the Phillips head screws from the bottom edge of the rear
panel, pull away the rear panel very carefully. Rough handling may cause
damage to the electrical connection for the fans.
1. Remove the (3) Phillips head screws from the bottom edge of the rear
panel.
2. Pull bottom edge 6” away from the cabinet, then pull the rear panel down
and away from the upper lip of the cabinet.
3. Unplug the fans from the centrifuge distribution PCB.
4. Remove the ground wire from the ground lug on the rear panel.
Note: Ensure that the EMC gasket material is not damaged or coming loose, if
so please replace. If this is not done the EMC of the device will be
compromised.
Fan Assembly
1. Remove the rear panel.
2. Remove the (2) nuts for each fan to be removed.
3. If removing less than all four fans, cut the wires so that they can be
reconnected by using crimp style connectors.
Cover Stops
1. Remove the rear panel.
2. Remove (2) Phillips head screws for each cover stop to be removed.