122
Calibration and Configuration
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
1. Access the Utilities mode of operation by pressing and holding
SHAKER
while switching the power on. Hold
SHAKER
until the UTLITIES menu
appears on the display.
2. Press
MODIFY PROGRAM
to move the arrow to CONFIGURATION and
then press
START
.
3. The message REMOVE MEMORY PROTECTION OR STRIKE STOP
appears.
4. Switch the memory protect switch located toward the bottom of the CPU
board (the left-most card cage board, looking from the rear of the unit) to
the Protect Off mode (down). The calibration screen appears with three
choices: PROTOCOLS CONFIGURATION, LANGUAGE, and
MACHINE.
5. Press
MODIFY PROGRAM
to move the arrow to MACHINE, then press
START
.
6. Press
MODIFY PROGRAM
until the arrow points to Data Logging and
then press
START
. The following screen is displayed:
7. Press
MODIFY PROGRAM
to move the arrow to DISABLE then press
START
. The following screen is displayed:
8. Press
STOP
three times. The following screen is displayed:
Data Logging
—> ENABLE
DISABLE
Status = XXXXX
Data Logging
ENABLE
—> DISABLE
Status = XXXXX
CONFIGURATION
PROTOCOLS CONFIGURATION
—> LANGUAGE
******************************
**
REPLACE MEMORY
**
**
PROTECTION
**
******************************