Troubleshooting
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
14
WATCHDOG ERROR
Safety system relay did
not open during system
start-up watchdog test.
1. Replace driver PCB
2. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
15
WATCHDOG ERROR
Safety system relay did
not open during system
start-up watchdog test.
1. Replace driver PCB.
2. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
16
WATCHDOG ERROR
Safety system relay did
not close during system
start-up watchdog test.
1. Replace driver PCB.
2. Replace CPU PCB.Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
17
WATCHDOG ERROR
Safety system did not
post fault to register
during system start-up
watchdog test.
1. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
Table 9, CPU error messages
Cod
e
Text Message
Description
Suggested Resolution