Disassembly
133
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Main Program
Assembly
1. Remove the processor PCB.
2. Using an antistatic IC removal tool for a 32 pin IC, gently pull the main
program IC from chip location U26. A future program upgrade may also
utilize U23.
Driver PCB
1. Remove the rear panel.
1. Unclasp the PCB levers and unplug the driver PCB from the backplane
PCB.
1. Slide the PCB out of the card cage.
Backplane PCB
1. Remove the (7) large Phillips head screws securing the backplane PCB
to the card cage frame.
2. Remove the (4) Phillips head screws securing the card cage frame to the
cabinet base (two in front, two behind).
3. Move the frame toward the rear of the cabinet, and rotate the base of the
backplane PCB out the front cavity of the cabinet.
4. Slide the frame out the back cavity of the cabinet.