Disassembly
129
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
General
Cautions
To avoid electrical shock, extreme caution should be exercised when working
inside the cabinet. Avoid contact with AC power connections and any
electrolytic capacitors, particularly on printed circuit PCBs.
Whenever possible, AC power should be disconnected from the device before
opening the cabinet.
Static
Discharge
The ACP 215 device uses static-sensitive electronics. Serious damage may
occur to the unit’s electronic systems if static discharge is not controlled.
Internal components should only be handled after proper precautions have
been taken to prevent static discharge. Service technicians must incorporate
the use of an anti-static service kit containing a wrist strap and anti-static mat.
Parts must only be placed on a properly grounded anti-static mat.
Notes
To reassemble the device, reverse the disassembly steps. In some cases,
reassembly notes have been added to direct your attention to any important
steps.
Label or tag each line as you disconnect it so you can easily identify where it
should be reconnected.
For brevity’s sake, a procedure will often reference another procedure. For
instance, the instructions for the removal of the top deck prompt you to first
remove the centrifuge. Locate the centrifuge removal procedure for instructions
on that particular step.
These instructions provide the most logical approach to disassembling the
device.