Installation:
Installation overview
3-2
Dash 3000/4000/5000
2000966-542D
Installation overview
The following illustrates the installation steps for patient-ready
monitoring.
1. Check received equipment against the sales order.
2. Unpack equipment and inventory contents.
3. Inspect the equipment and its components carefully prior to
installation. Refer to
4. (Optional) Insert two batteries into the battery compartment slots.
(Batteries should be charged for 4 hours before using device on
patients.)
5. Connect the power cord to the power supply inlet on the Dash patient
monitor. Secure the power cord to the back of the Dash using the
power cord retention clamp, unless the Dash is used with the Dash
Port 2 docking station. Then plug the power cord into the AC power
outlet.
6. Connect peripheral equipment and network, if applicable.
7. Power up the Dash monitor. Go to
on page 4-1 to
configure for stand-alone or networked monitoring.
8. Connect patient cables per the operator’s manual.
9. Qualified personnel must perform the
on page 3-8, and should record the values of each required
599A
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......