Equipment overview:
Components
2-8
Dash 3000/4000/5000
2000966-542D
even though they will be charged sequentially. The battery icon lights
green when the respective battery is fully charged.
When the patient monitor is operating under battery power the battery
icons are not illuminated. The icons are also not illuminated when the
respective battery is either not being charged, not installed, or has failed.
The following table explains what the charging status indicators mean.
NOTE
No specific indicator distinguishes a failed battery pack condition
from a condition where the battery is not installed or is not being
charged. Go to the
Service Menu
for
Battery Status
. Refer to
on page 6-7 for further information.
Battery status indicators
The battery status indicators are located inside the battery
compartment. One green LED indicator is located above each of the two
battery slots and lights green when the patient monitor is receiving
power solely from the respective battery. The indicators do not
illuminate when the patient monitor is not battery powered.
Neither indicator lights when the patient monitor is operating from both
batteries simultaneously (e.g., in a very low battery charge condition
when both batteries are joined together in order to sustain operation of
the patient monitor).
Battery capacity gauge
On-screen capacity gauges indicate the battery’s current state of health
and charge status. A battery capacity gauge for each battery present
displays below the parameter blocks in the lower right corner of the
display. The capacity gauge indicates the remaining charge capacity
(usable energy left) for each battery.
LED color
Explanation
Yellow
Two battery icons, labeled
Charging Status A
and
B
, illuminate yellow
when the respective battery is being charged. If both batteries are present
and require charging, then both icons illuminate yellow even though they
charge sequentially.
Green
The icon lights green when the respective battery is fully charged.
No light
The icon does not illuminate under the following conditions:
The respective battery is not installed.
The patient monitor is operating on battery power.
A failure condition has been detected for the respective battery.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......