Configuration:
Service menus
4-6
Dash 3000/4000/5000
2000966-542D
Service Mode options
Following is a list of options in the main code service menu.
Service Mode menu options
Menu option
Description
REVIEW ERRORS
For advanced troubleshooting by GE engineers. Error
log data can be transferred over the network to a central
station and then loaded onto a diskette for review. Refer
to
“Error messages”
on page 6-5 for details.
CALIBRATE
For checkout or calibration of the noninvasive blood
pressure, ECG analog output, BP analog output, CO
2
service, and SAM service menu functions of the patient
monitor.
BATTERY SERVICE
Provides battery data.
NOTE
If an
INCOMPAT
message displays in the Manufacturer Name
column and
UNKNOWN
displays in all other columns, the
installed battery is incompatible.
PATIENT-MONITOR TYPE
For selecting the type of patient monitor desired, i.e
adult, neonatal or operating room. Refer to
on page 4-12 for detailed procedures.
WARNING
Changing the patient-monitor type will default the
admit function to Standard configuration. Different
alarms and parameters are activated for each selection.
NOTE
The keypad/remote control is DIDCA programmed for specific
patient monitor types. The error message, “
WARNING:
REMOTE MISMATCHED WITH MONITORING MODE
”
displays if the patient monitor and keypad/remote control do not
match.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......