Equipment overview:
Theory of operation
2000966-542D
Dash 3000/4000/5000
2-37
Single battery run time: approximately 2.5 hours –
An icon fully
outlined with solid lines and completely filled indicates a new battery,
fully charged to its Design Capacity.
Single battery run time: approximately 2.0 hours –
The dotted
portion of the icon outline shows that the battery has lost 20% of its
Design Capacity. The solid-outline portion is filled in, showing that the
battery is charged to 100% Full Charge Capacity.
Single Battery Run Time: Approximately 1.0 Hours –
The dotted
portion of the icon outline shows that the battery has lost 20% of its
Design Capacity. The solid-outline portion is half filled in, showing that
the battery is charged to only 50% of available capacity.
Identify battery capacity using the Cadex SMart Two+
charger
NOTE
The Cadex SMart Two+ charger is no longer available for purchase.
However, if your site has a Cadex SMart Two+ charger, you should
continue to use it to charge and condition the battery.
When you select the
Target Capacity
switch on the charger, the
charger compares the battery’s performance to a 60%, 70%, or 80% target
capacity set on the battery charger. If the battery fails to meet the target
performance, the battery charger will prompt you to condition the
battery. If after the condition cycle has been completed, the battery does
NOT meet the target performance, then the battery charger will
illuminate a “fail” light.
857A
Design Capacity and Full Charge Capacity
858A
Full Charge Capacity
859A
Full Charge Capacity
50% Available Capacity
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......