Functional and electrical safety checks:
Functional Checkout procedures
8-44
Dash 3000/4000/5000
2000966-542D
1. Select
MORE MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
3. Select
CALIBRATE
>
CAL BP ANALOG OUT
>
SET BP LOW
.
4. Adjust the count for –9.0 V ± 5 mV on the meter and press the
Trim
Knob
control.
5. Select
SET BP HIGH
.
6. Adjust the count for +9.0 V ± 5 mV on the meter and press the
Trim
Knob
control.
7. Select
SET BP ZERO
. Adjust the count for 0.0 V ± 5 mV on the
meter and press the
Trim Knob
control.
8. Select
CONFIRM BP CAL
to confirm or abort the calibration.
End-tidal CO
2
test (option)
To verify the mainstream end-tidal CO2, refer to the CO2 chapter in the
Dash 3000/4000/5000 Patient Monitor Operator’s Manual. This test
requires you perform a zero and reference check by using the sample
cells provided on the end-tidal CO2 cable.
For instructions on verifying sidestream end-tidal CO2 and setting the
sample line to zero, refer to the CO2 module operator instructions.
Battery tests
1. Check battery power for both batteries.
Unplug the patient monitor (or the docking station) from AC
power and open the battery door. Verify one LED in the battery
compartment is on (batteries must have more than 10% charge).
Pull that battery out and verify the other LED lights, thus
indicating the unit is powered by the other battery.
Reinstall battery and plug in patient monitor to a wall outlet (or
the docking station).
Graph or print tests (option)
Communication confirmation
Confirm communication across the network as follows.
1. Admit and generate a waveform at the patient monitor with a
simulator.
2. Press
and observe graph output at chosen locations.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......