Functional and electrical safety checks:
Functional Checkout procedures
2000966-542D
Dash 3000/4000/5000
8-25
Respiration tests
Connect the ECG patient cable to the
ECG/RESP
connector of the
patient monitor.
1. Set up the patient simulator as follows:
Respiration (RESP) baseline impedance – 750
or 1000
RESP
R – 0.5
Select appropriate lead for RESP on simulator
RESP rate (respirations per minute) – 30
2. Set up the patient monitor as follows:
RESP waveform – on
RESP waveform lead select – lead II (RESP waveform derived
from ECG lead II)
3. Observe the following:
RESP parameter window appears on the patient monitor with a
reading of 30 ±2 (respirations per minute)
RESP waveform appears distortion-free on the patient monitor
4. Change the RESP waveform lead select of the patient monitor to lead
I (RESP waveform derived from ECG lead I), LA at the simulator.
5. Observe the following:
RESP parameter window appears on the patient monitor with a
reading of 30 ±2 (respirations per minute),
RESP waveform appears distortion-free on the patient monitor.
6. Disconnect the ECG patient cable from the
ECG/RESP
connector of
the patient monitor.
7. Proceed to the next steps in these checkout procedures.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......