Field replaceable units:
Replace display assembly parts
7-32
Dash 3000/4000/5000
2000966-542D
1. Open the display assembly. Refer to page
.
2. Remove the LCD display. Follow the steps for the patient monitor
you are servicing.
Dash 3000 patient monitor
with display shield flex circuit
Dash 4000 patient monitor
Dash 5000 patient monitor
a. Peel back and remove the rubber
display isolator from around the
display.
b. Go to next step.
a. Disconnect the five cable connectors.
b. Remove the four screws anchoring
the mounting plate to the LCD display.
c.
Remove the mounting plate and set
aside.
a. Disconnect the four or five cable
connectors, as required.
b. Remove the mounting plate and set
aside.
c.
Lift out the LCD display from the
rubber display isolator and set aside.
504A
885A
885A
8851A
8871A
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......