Functional and electrical safety checks:
Functional Checkout procedures
8-46
Dash 3000/4000/5000
2000966-542D
Speaker test
1. Change the alarm volume of the patient monitor to 100%.
2. Verify the speaker volume of the patient monitor changes
accordingly.
3. Return the volume of the patient monitor to the level it was
previously set to, before you changed it for this test.
Network test (option)
1. Verify that the patient monitor is connected to the CARESCAPE
Network MC (Mission Critical).
NOTE
When the patient monitor is connected to a docking station,
verify the docking station is connected to the CARESCAPE
Network MC.
NOTE
When the patient monitor is connected to a docking station, the
patient monitor’s network port is disabled. Only the docking
station’s network port is enabled at this time.
2. Select
VIEW OTHER PATIENTS
.
3. Select
SELECT ANOTHER CARE UNIT
.
4. Verify that you can see at least one care unit.
5. Select a care unit.
6. Select
SELECT A BED TO VIEW
.
7. Select a bed.
8. Verify that the patient window appears on the patient monitor’s
split-screen.
Remote control test (option)
1. Verify the remote control is connected into an Autoport to M-Port
adapter and is inserted into the
Aux
connector.
2. Select
MORE MENUS
>
MONITOR SETUP
>
REVISION AND ID
.
3. Select
NEXT
from the popup menu to display the port connectors.
4. Verify the Remote Control label appears after the appropriate port
and the software version for the remote control is shown.
5. Press each remote control key and verify a beep tone sounds at the
patient monitor.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......