Functional and electrical safety checks:
Functional Checkout procedures
8-34
Dash 3000/4000/5000
2000966-542D
5. Verify the following SpO2 readings on the Dash for saturation and
pulse rate:
Saturation (%): 75 ±2
Rate (bpm): 60 ± 2
6. Press and release the
HEART RATE
switch and verify the SRC-
MAX default indicators are lit as follows:
HEART RATE = 200 bpm
LIGHT = LOW
%SpO2 = 75
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.
7. Verify the following SpO2 readings on the Dash for saturation and
pulse rate:
Saturation (%): 75 ±2
Rate (bpm): 200 ±3% (194 to 206)
8. Press and release the
LIGHT
switch and verify the SRC-MAX
default indicators are lit as follows:
HEART RATE = 200 bpm
LIGHT = HIGH
%SpO2 = 75
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.
9. Verify the following SpO2 readings on the Dash for saturation and
pulse rate:
Saturation (%): 75 ±2
Rate (bpm): 200 ±3% (194 to 206)
10. Press and release the %SpO2 switch and verify the SRC-MAX
default indicators are lit as follows:
HEART RATE = 200 bpm
LIGHT = HIGH
%SpO2 = 90
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.
11. Verify the following SpO2 readings on the Dash for saturation and
pulse rate:
Saturation (%): 90 ±2
Rate (bpm): 200 ±3% (194 to 206)
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......