Electromagnetic compatibility (EMC):
Electromagnetic compatibility (EMC)
2000966-542D
Dash 3000/4000/5000
A-3
Guidance and manufacturer’s declaration – electromagnetic
immunity
The Dash 3000/4000/5000 is intended for use in the electromagnetic
environment specified in the following table. It is the responsibility of the
customer or user to assure that the Dash 3000/4000/5000 is used in such
an environment.
CAUTION
There are technological limitations in the ability of the
monitor to perform as intended when subjected to
interference from electrostatic discharges (ESD) at or
above ±8kV when tested to EN 60601-1-2. ESD air
discharges into the patient cable connectors and the
Aux
port may cause the monitor to reboot and recover to its
normal operating mode without operator intervention.
The following guidelines can greatly reduce the potential
for ESD interference prior to connecting any cables to the
monitor:
Discharge any electrostatic charge from your body
prior to connecting any cables to the monitor.
You can discharge electrostatic charge by touching
the exposed metal equipotential terminal on the back
of the monitor. Refer to
on page 2-4 for
location of the equipotential terminal.
NOTE
ESD will not cause any permanent damage to the monitor.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......