Troubleshooting:
Get error logs
2000966-542D
Dash 3000/4000/5000
6-17
Connect the PC to Dash patient monitor or MC network
NOTE
Connect the patient monitor and PC to a reliable power source
during the transfer process. Do
not
use battery power. A battery
failure during the transfer process may result in a transfer process
failure.
1. Connect one end of the category 5 crossover or network patch cable to
the PC network card.
2. Connect the other end of the category 5 crossover patch cable as
follows:
Connect to the
ETHERNET
port of the docking station if the
patient monitor is using a Dash Port 2 docking station.
Connect to the
ETHERNET
port of the patient monitor if no
docking station or network is used.
Connect to the MC network Ethernet tap if transferring logs
from other devises on the network. If you connect directly to the
MC network Ethernet tap, verify that the patient monitor or
docking station is also connected to the MC network Ethernet
tap.
3. Power on the PC.
4. Continue with the instructions,
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......