Functional and electrical safety checks:
Functional Checkout procedures
2000966-542D
Dash 3000/4000/5000
8-37
mmHg) for at least one full minute. If the readings are not equal for
at least one full minute, the NBP circuit requires calibration. See
4. Select
CALIBRATE
>
CALIBRATE NBP
>
CHECK CAL OFF
>
START
.
5. Remove the NBP test setup apparatus from the patient monitor. The
NBP tests are complete.
NBP calibration
NOTE
Only perform this test if the NBP pressure test fails.
The overall accuracy of noninvasive blood pressure (NBP) readings by
the patient monitor depend on the following:
the zero pressure reading, and
the voltage span of the NBP sensor in the patient monitor.
This procedure provides a method of verifying these items are accurate
and also checks the NBP pneumatic circuit plumbing for leaks.
WARNING
When the NBP cuff is used in this procedure, it must be
tightly wrapped around a rigid cylinder or pipe.
Do not
put the NBP cuff around a human arm during the
calibration procedures due to the potential for injury.
Calibration procedure
Using the
Trim Knob
control, access the
Service Mode
menu starting
from the Main Menu.
1. Select
MORE MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros (e.g. July 4 =
0407).
3. Select
CALIBRATE
>
CALIBRATE NBP
>
CAL ZERO OFF
>
START
.
4. The text on the menu item changes from Cal Zero Off to Cal Zero In
Progress.
When the process is complete, the menu item shows that it is
OFF
again.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......