Troubleshooting:
Fault analysis
6-4
Dash 3000/4000/5000
2000966-542D
Acquisition PCB symptoms
Symptoms relative to patient signal acquisition such as missing
parameter text and waveform(s) may be associated with acquisition PCB
failure. It is important that you are able to distinguish the difference
between the general format of the display, which is generated by the
processor PCB, versus the patient signals and data that is associated
with these patient signals, a function of data acquisition, which is
generated by the acquisition PCB.
Processor PCB symptoms
Symptoms with battery operation, network communications,
asynchronous communications, NBP control, PCMCIA communications,
analog output, audio/sound generation, and video signals as well as other
display-related problems all may be associated with processor PCB
failure. All of these are functions controlled by microcontroller or
graphics processing circuitry located on the Processor/Power
Management PCB.
Network connector
No network connection when a
network cable is connected to the
patient monitor’s Network Connector.
The Dash patient monitor is
connected to a Dash Port 2 docking
station. The patient monitor’s
Network Connector is inactive when
it is connected to the docking
station.
Network is disabled in boot code.
Connect the network cable to the docking station’s
internet connector.
Enable network in boot code.
The Dash patient monitor does not
have the optional CARESCAPE
Network option enabled.
Purchase and enable the CARESCAPE Network
option.
Remote control
The remote control does not activate
the display.
Cable or Autoport to Mport adapter
connections may be loose.
Restore the connection.
Remote control needs to be reset.
Disconnect and securely reconnect the remote
control from the patient monitor.
Remote control is defective.
Replace the remote control with a known good one.
Masimo SET or Nellcor SPO2
Probe or module malfunction
Five or more consecutive hardware
failures have occurred.
Review the input error log for Masimo or Nellcor
failure codes.
Problem
Possible reason
Solution
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......