Equipment overview:
Theory of operation
2-22
Dash 3000/4000/5000
2000966-542D
Cardiac output
The cardiac output function measures blood temperature and injectate
temperature, and uses the thermal dilution method to calculate cardiac
output. Patient alarms with adjustable high and low limits for blood
temperature are provided. System alarms for sensor failure and unstable
blood temperature are provided.
The patient monitor accepts the brown color-coded connector (compatible
with the Eagle 4000 monitor and Tram modules).
The cardiac output connector and measurement circuits are shared with
the temperature monitoring function. You cannot use both functions
concurrently. A signal in the patient cable indicates the appropriate
function.
Carbon dioxide (CO
2
)
The CO
2
function measures inspired and expired CO
2
and respiration
rate using the infrared light absorption technique. The patient monitor
connects to an external CapnoFlex Low Flow Sidestream module or
Novametrix Capnostat III sensor that clips to an airway adapter in the
patient’s ventilation circuit. The circuits to drive the sensor and process
its incoming signal are located within the DAS.
Patient alarms with adjustable high and low limits for inspired CO
2
,
expired CO
2
, and respiration rate are provided. An additional patient
alarm for no breath detected is provided. System alarms for various
sensor conditions are provided.
The patient monitor accepts the yellow color-coded connector.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......