Functional and electrical safety checks:
Functional Checkout procedures
8-20
Dash 3000/4000/5000
2000966-542D
Functional Checkout procedures
Frequency
Qualified personnel must perform the checkout procedures:
Every 12 months after receipt of the device (Preventative
Maintenance). Refer to
on page 5-2 for more
information.
Each time the main enclosure is disassembled or a circuit board is
removed, tested, or replaced (Corrective Maintenance). Refer to
on page 7-52 for more information.
Identify enabled patient parameters and software options
The check out procedures support all enabled patient parameter and
software options. To view the options enabled on the patient monitor, go
to the Main Menu and select
MORE MENUS
>
MONITOR SETUP
>
SOFTWARE CONFIGURATION
.
The following is a list of all available patient parameter and software
options.
0BP (invasive)
1BP or 2BP (invasive)
3BP or 4BP (invasive)
12SL
ACI-TIPI
HI-RES TRENDS
ETCO
2
(CO
2
) (End-tidal CO
2
)
NETWORK
AVOA Plus
CARDIO-PULMONARY
CARDIAC
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......