Functional and electrical safety checks:
Functional Checkout procedures
8-24
Dash 3000/4000/5000
2000966-542D
12SL and ACI-TIPI ECG test (option)
Perform this test if your patient monitor uses the 12SL ACI-TIPI ECG
option.
1. Set up the patient simulator as follows:
Heart rate – 80 bpm
Heart rate amplitude – 1.0 mV
12SL ECG patient cable (5-leadwires with V leadwires) properly
attached.
2. Select
ECG
from the patient monitor menu. Then, select
12 LEAD
ECG ANALYSIS
.
3. Verify that the patient monitor is displaying 12 noise-free leads.
4. Select
12LD ECG NOW
. Wait for the patient monitor to acquire and
analyze the data.
5. Select Transmit-Print.
6. Verify the 12SL ECG prints at the print location assigned in the
patient monitor’s
Print Setup
>
12SL Print Location
.
If there is no print location is assigned, an error message appears
on the bottom of the patient monitor’s display.
7. Verify the ECG is transmitted to the MUSE Cardiovascular
Information System. Verify the ECG prints out correctly as defined
by the MUSE system.
If no MUSE system is connected, an error message appears on
the bottom of the patient monitor’s display.
8. Delete this test 12SL ECG from the MUSE system’s edit list.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......