Functional and electrical safety checks:
Functional Checkout procedures
8-40
Dash 3000/4000/5000
2000966-542D
Analog output and defibrillator synchronization tests
Equipment
Use the following equipment for these tests:
Oscilloscope, or equivalent
5.5 digit precision voltmeter (HP34401A or equivalent). Needed for
calibration only.
Analog output cable (2000633-001 or equivalent)
Setup
1. Use the figure below as a reference for connecting test devices to the
DEFIB SYNC
connector, located on the back panel of the patient
monitor, for performing these tests.
2. Test the ECG, Arterial BP, and Marker Out signals from the
DEFIB
SYNC
connector. ECG Input voltage of 1 mV = 1 V analog -
Analog output cable pinout (2000633-001)
PIN number
Color
1
Black
2
Green
3
Red
4
Brown
5
Blue
6
White
7
Yellow
8
Gray
039B
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......