Troubleshooting:
Monitor defaults transfer
2000966-542D
Dash 3000/4000/5000
6-11
Monitor defaults transfer
Below is a list of error messages that may display when performing a
patient monitor defaults transfer.
Storing monitor defaults
Copying stored monitor defaults
Message
Description
ERROR - UNIT DEFAULTS HAVE NOT
BEEN SAVED
Patient monitor defaults were not saved on
the server patient monitor due to a memory
error in the patient monitor.
Message
Description
ERROR COPYING UNIT DEFAULTS -
NETWORK ERROR
Either an error has occurred while the
defaults were being transferred, or defaults
have not been saved on the server patient
monitor.
ERROR COPYING UNIT DEFAULTS -
UNIT TYPE MISMATCH
The server and client patient monitors are
set to different unit types (e.g.,
ADULT
,
NEO
, or
OR
)
ERROR COPYING UNIT DEFAULTS -
SOFTWARE VERSION MISMATCH
The server and client patient monitors have
different software versions (e.g., V2C, V3A).
ERROR COPYING UNIT DEFAULTS -
DATA CORRUPTION ERROR
The patient monitor defaults transferred have
become corrupted, possibly due to a
memory error.
ERROR COPYING UNIT DEFAULTS -
SOFTWARE FEATURE LEVEL
MISMATCH
The server and client patient monitors are
set to different software levels (e.g., Basic,
Cardiac, Cardiopulmonary).
ERROR COPYING UNIT DEFAULTS -
COUNTRY CODE MISMATCH
The server and client patient monitors are
set to different country codes (e.g.,
DEFAULT
or
FRANCE
).
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......