Field replaceable units:
Recommended checkout
7-52
Dash 3000/4000/5000
2000966-542D
Recommended checkout
After reassembling the patient monitor, ALWAYS complete the electrical
safety tests, checkout procedures, calibration tests, and regular
maintenance procedures identified and described in this section. Safety
and functional checkout tests should be documented in the checklist
provided in
Recommended calibration, electrical safety tests, and checkout procedures
Replacement
procedure
Calibration
tests
Checkout procedures
Electrical safety tests
Handle assembly
None
“Patient monitor power-up tests” on page 8-21
“Wireless antenna signal strength test (option)”
on page 8-49
. Use this procedure if the wireless
option is installed.
“Wireless LAN test (option)” on page 8-52
this procedure if the wireless option is installed.
“Electrical safety tests” on page 8-4
Display assembly
Display assembly
components
Display inverter
assembly
None
“Patient monitor power-up tests” on page 8-21
Speaker
None
“Patient monitor power-up tests” on page 8-21
“Wireless antenna signal strength test (option)”
on page 8-49
. Use this procedure if the wireless
option is installed.
“Wireless LAN test (option)” on page 8-52
this procedure if the wireless option is installed.
Keypad assembly/
Trim Knob
None
“Patient monitor power-up tests” on page 8-21
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......