2D
Measurement
Accuracy and
Range
System
Tolerance
a
Accuracy By
Test Method
Range
(vertical)
of full scale
Linear distance
(horizontal)
˂± 2% plus 1%
of full scale
Acquisition
Phantom
0-32 mm
Traced area
˂± 3% plus 1%
of full scale
Acquisition
Phantom
0-1206.2 mm
2
Traced
distance
˂± 2% plus 1%
of full scale
Acquisition
Phantom
0-139 mm
a. Full scale for distance implies the maximum depth of the image.
2D
Measurement
Accuracy and
Range
System
Tolerance
Accuracy By
Test Method
Range
Distance
˂± 2% plus 1%
of full scale
a
Acquisition
Phantom
0.35.5 mm
Time
˂± 0.2% plus
1% of full
scale
b
Acquisition
Phantom
0.30 sec
a. Full scale for distance implies the maximum depth of the image.
b. Full scale for time implies the total time displayed on the scrolling graphic image.
M-Mode Measurement Accuracy and Range
Sources of measurement errors
In general, two types of errors can be introduced into the measurement:
408
Appendices