Direct, indirect, and receive controls
Direct Controls
The system does not have a direct user control for output. Rather, the system has
been designed to automatically adjust output to ensure that acoustic and thermal
limits are not exceeded for all imaging modes. Since there is no direct user control for
output, the sonographer should rely on controlling exposure time and scanning
technique to implement the ALARA principle.
The system does not exceed a spatial peak temporal average intensity (ISPTA) of
720 mW/cm2 for all imaging modes. The mechanical index (MI) and thermal index (TI)
may exceed values greater than 1.0 on some transducers in some imaging modes.
Ultrasound users can monitor the MI and TI values on the left side of the clinical
monitor under
Transmit
and implement the ALARA principle accordingly. For more
information on MI and TI, see
Indirect Controls
The controls that indirectly affect output are controls affecting imaging mode, freeze,
and depth. The imaging mode determines the nature of the ultrasound beam. Freeze
stops all ultrasound output but keeps the last image displayed on screen. Freeze can
be used by the ultrasound user to limit exposure time while studying an image and
maintaining probe position during a scan. Some controls, such as depth, show a rough
correspondence with output, and may be used as a general means for indirectly
reducing MI or TI. See
"Guidelines for reducing MI and TI"
Receive Controls
The receive controls are the gain controls. Receive controls do not affect output. They
should be used, if possible, to improve image quality before using controls that
indirectly affect output.
350
Appendices
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...