WARNING:
To avoid the risk of injury, do not operate the system in the
presence of flammable gasses or anesthetics. Explosion can result.
WARNING:
The power frequency magnetic field should be measured in the
intended installation location to assure that it is sufficiently low. If image
distortion occurs, it may be necessary to position the Vevo MD Imaging
System further from sources of power frequency magnetic fields.
WARNING:
Because the only method of completely removing AC power
from the system is to disconnect the AC input power cord from the stand
base, ensure that you place the system in a location in which you can easily
remove the AC input power cord if necessary.
Caution:
Avoid increasing the system and transducer connector temperature, do
not block the airflow to the front and back of the system.
Moving and setting up the system
21