Immunity
test
IEC 60601 test
level
Compliance level
Electromagnetic
Environment
Power
Frequency
Magnetic
Field IEC-
61000-4-8
3 A/m
3 A/m
If image distortion
occurs, it may be
necessary to position
the Vevo MD Imaging
System further from
sources of power
frequency magnetic
fields or to install
magnetic shielding. The
power frequency
magnetic field should be
measured in the
intended installation
location to assure that it
is sufficiently low.
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the Vevo MD
Imaging System
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
Separation Distance
d
=1.2
√P
Radiated RF
3 Vim
3 V/m
d
=1.2
√P
Manufacturer's declaration