Quantity
Precision
(% of standard deviation)
Uncertainty (%)
(95% confidence)
fc
1.1
±3.0
PII
14.9
+107.8 to -112.7
PII
.3
15.5
+109.6 to -114.0
Quantity
Precision
(% of standard deviation)
Uncertainty (%)
(95% confidence)
P
r
26.5
+60.5 to -72.8
P
r.3
27.5
+60.9 to -73.1
Wo
18.2
+118.8 to -144.0
fc
1.1
±3.0
PII
14.9
+118.6 to -143.9
PII
.3
15.5
+119.6 to -144.4
Acoustic measurement precision and uncertainty for UHF70
Factors that contribute to display uncertainty
The net uncertainty of the displayed indices is derived by combining the quantified
uncertainty from three sources: measurement uncertainty, system and transducer
variability, and engineering assumptions and approximations made when calculating
the display values.
Measurement errors of the acoustic parameters when taking the reference data are
the major source of error that contributes to the display uncertainty. The measurement
error is described in
"Acoustic measurement precision and uncertainty"
Acoustic measurement precision and uncertainty
401
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...