10 Clean
the system to remove any visible gel, debris or bodily fluids.
♦
Wipe
from clean areas to the soiled areas to avoid cross-contamination.
11 Verify
that all gel, debris, and bodily fluids have been removed from the transducer
and system. If necessary, repeat Steps 9 and 10, with a new wipe.
12 Obtain
a compatible disinfectant. Compatible disinfectant materials listed in the
table
"Compatible cleaners and disinfectants"
Caution:
Make sure the solution does not get into any of the connector ports.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces if not using a wipe.
13 Disinfect
the system by wiping all surfaces including:
l
clinical display (upper monitor)
l
control panel (lower monitor)
l
transducer and gel holders
l
cable hooks
l
grab bars
l
all surfaces that may have been touched and/or contaminated
14 Ensure
the system stays wet with the disinfectant for the wet contact time listed
in the table
"Compatible cleaners and disinfectants"
This may require
reapplication to achieve the required wet time
.
♦ For example, if you are using Sani-Cloth Bleach it has a wet contact time of 4
minutes. Wipe the entire system until thoroughly wet. If the system starts to dry
before the 4 minutes is reached, use a new wipe and re-wet the system. Continue the
process of re-wetting with a new wipe until a total wet time of 4 minutes is reached.
15 Allow
the system to air dry in a clean, well ventilated space. Do not dry with a
cloth.
322
Cleaning and disinfecting
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...