Step A: Semi-critical contact cleaning and disinfecting of
system and transducer
Part 1 of 2: Cleaning of system and transducer, disinfecting of system for
semi-critical contact
Follow the manufacturer's instructions when using cleaners and disinfectants. The
materials listed in the table below are chemically compatible with the system and
transducer. Confirm that cleaning and disinfecting materials are appropriate for your
facility’s use.
Product
System
Transducers
(UHF22, UHF48, UHF70)
Wet Contact Time
Tristel Sporicidal Wipes
30 seconds
Microbac Tissues
1 minutes
Cleanisept wipes
2 minutes
Sani-Cloth AF3 (grey
top)
*
3 minutes
Sani-Cloth Bleach
(orange top)
*
4 minutes
Sani-Cloth HB
(green top)
10 minutes
Compatible cleaners and disinfectants
Step A: Semi-critical contact cleaning and disinfecting of system and
transducer