
95
10 Basic Measurement
• Place the mounting shoe on the pipe. Hold clasp and mounting shoe with one hand
while pushing the tension strap through parts (2) and (1) of the clasp (see Fig. 10.11).
• Pull the tension strap firmly and engage it in the inner hooks of the clasp.
• Mount the second transducer in the same way. Adjust the displayed transducer dis-
tance by means of the measuring tape.
• Tighten the screws of the clasps.
• Push the transducers in the mounting shoes. Press the transducer firmly on the pipe.
There should be no air pockets between transducer surface and pipe wall. Tighten the
screw of the mounting shoe firmly.
Note!
The clasp and the metal spring must be in contact completely with
the pipe to ensure a good fixation.
Note!
At pipes with large diameters, use tongs, if necessary, to tense the
tension strap.
Note!
When the transducers are mounted on a vertical pipe and the flow-
meter is placed lower than the pipe, the cable of the upper transducer
should be fixed to the tension strap by a cable tie to protect the cable
from mechanical strain.
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......