204
Technical Data
Technical Data
Damping Mats (Option)
Damping mats will be used for the gas measurement to reduce noise influences on the
measurement.
Transducer damping mats
will be installed below the transducers.
Pipe damping mats
will be installed at reflection points, e.g. flange, welding.
Selection of Damping Mats
type
description
outer pipe
diameter
dimensions
l x b x h
transducer
frequency
tempera-
ture
remark
mm
mm
G H K M P
°C
transducer damping mat
C
self-adhesive, for
stationary installati-
on
< 80
450 x 115 x 0.5
-
-
-
x
x
-25...+60
80
900 x 230 x 0.5
-
-
x
x
-
900 x 230 x 1.3
x
x
-
-
-
pipe damping mat
B
self-adhesive, for
stationary installati-
on
l x 100 x 0.9
x
x
x
x
x
-35...+50 l - see
table
below
Pipe Damping Mat Type B: Length l Depending on Transducer Fre-
quency and Outer Pipe Diameter
outer pipe diameter
transducer frequency
mm
G, H
K, M, P
100
2 m
1 m
200
6 m
3 m
300
12 m
6 m
500
32 m
16 m
1000
126 m
63 m
b
l
D
transducer damping mat
pipe damping mat
reflection mode
diagonal mode
D - outer pipe diameter
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......