162
18 Outputs
18.6 Behavior of the Alarm Outputs
18.6.1 Apparent Switching Delay
Measured values and totalizer values will be displayed rounded to two decimal places.
The limits, however, will be compared to the non-rounded measured values. This might
cause an apparent switching delay when the measured value changes marginally (less
than two decimal places). In this case, the switching accuracy of the output is greater
than the accuracy of the display.
18.6.2 Reset and Initialization of the Alarms
After a cold start, all alarm outputs will be initialized as follows:
Press three times key C during measurement to set all alarm outputs to the idle state.
Alarm outputs whose switching condition is still met will be activated again after 1 s. This
function is used to reset alarm outputs of type
HOLD
if the switching conditions is not met
anymore.
By pressing key BRK, the measurement will be stopped and the main menu selected. All
alarm outputs will be de-energized, independently of the programmed idle state.
18.6.3 Alarm Outputs during Transducer Positioning
When the positioning of the transducers begins (bar graph display), all alarm outputs
switch back to the programmed idle state.
If the bar graph is selected during measurement, all alarm outputs switch back to the pro-
grammed idle state.
An alarm output of type
HOLD
being activated during the previous measurement remains
in the idle state after transducer positioning if the switching condition is not met anymore.
Switching of the alarms into the idle state will not displayed.
18.6.4 Alarm Outputs during Measurement
An alarm output with switching condition
MAX
or
MIN
will be updated max. once per sec-
ond to avoid humming (i.e. fluctuation of the measured values around the value of the
switching condition).
An alarm output of type
NON-HOLD
will be activated if the switching condition is met. It
will be deactivated if the switching condition is not met anymore. The alarm remains acti-
vated min. 1 s even if the switching condition is met shorter.
Alarm outputs with switching condition
QUANTITY
will be activated when the limit is
reached.
Table 18.7: Alarm state after initialization
FUNC
OFF
TYPE
NON-HOLD
MODE
NO CONT.
LIMIT
0.00
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......