123
13 Storing and Output of Measured Values
13.2.5 Data Format
The header will be transmitted at the beginning of the measurement. The first four lines
contain general information about the flowmeter and the measurement. The following
lines contain the configuration parameters output for each channel in a data block.
The line
\DATA
will be transmitted next, followed once by the column titles (see Table
13.1) for the corresponding channel. The measured values will be transmitted after-
wards.
example:
\DEVICE :G70X-XXXXXXXX
\MODE :ONLINE
\CHAN :1 (A:)
DATE :09.01.2008
TIME :19:56:52
Par.Record
Meas. Point No. :A:F5050
Pipe
Outer Diameter :60.3 mm
Wall Thickness :5.5 mm
Roughness :0.1 mm
Pipe Material :Carbon Steel
Lining :NO LINING
Medium :natural gas
Medium Temperature :38 C
Medium pressure :60.00 bar
Transducer Type :xxx
Sound Path :3 NUM
Transducer Distance :-15.6 mm
Damping :20 s
Full-Scale Val. :4.50 m3/h
Physic. Quant. :volume flow
Unit of Measurement :[m3/h]/[m3]
example:
\DATA
A:;\*MEASURE; Q_POS; Q_NEG;
B:;\*MEASURE; Q_POS; Q_NEG;
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......