210
B Menu Structure
input of the kinematic viscosity of the medium
range: 0.01...30 000 mm
2
/s
This display is indicated only if
OTHER ME-
DIUM
or
NATURAL GAS
has been selected.
input of the operating density of the medium
range: 0.1...20 000 kg/m
3
, if
SPECIAL
FUNCTION\SYSTEM SETTINGS\MEASUR-
ING\GAS-MEASURING
is activated or
0.01...20 g/cm
3
, if
SPECIAL FUNC-
TION\SYSTEM SETTINGS\MEASURING\
GAS-MEASURING
is deactivated
This display is indicated only if
OTHER ME-
DIUM
or
NATURAL GAS
has been selected.
input of the gas compressibility factor
range: 0.001...2
This display is indicated only if
OTHER ME-
DIUM
or
NATURAL GAS
has been selected.
input of the medium temperature
default: 20 °C
input of the medium pressure
range: 1...600 bar
This display is indicated only if
GAS-MEA-
SURING
in
SPECIAL FUNCTIONS\SYSTEM
SETTINGS\MEASURING
is activated or if the
GAS-MEASURING
is deactivated and
FLUID
PRESSURE
in
SPECIAL FUNCTIONS\SYS-
TEM SETTINGS\DIALOGS/MENUS
has been
activated.
selection of the transducer type
This display will be indicated only if no or spe-
cial transducers are connected.
cold start
resistant
kinem. Viscosity
1.00
mm2/s
Density
60.00
kg/m3
Gas compr.factor
1.000
factor
Medium Temperat.
20.0
C
Fluid Pressure
60.00
bar
Transducer Type
Standard
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......