6 Installation of FLUXUS G704
6.4
Connection of the Transducers - Connection System TS
It is recommended to run the cables from the measuring point to the flowmeter before the
connection of the transducers to avoid load on the connection points.
Overview
6.4.1 Transducers (ATEX Zone 1) - Direct Connection
Observe the safety instructions in chapter 2.
• Remove the left blind plug for the connection of the transducers (see Fig. 6.3).
• Open the cable gland of the transducer cable. The compression part remains in the
cap nut.
• Push both transducer cables through cap nut and compression part.
• Prepare the transducer cables with the cable gland. Cut the outer shield and brush it
back.
• Tighten the gasket ring side of the basic part in the housing.
• Insert the transducer cable in the housing.
• Fix the cable gland by screwing the cap nut on the basic part.
• Connect the leads to the terminals of the flowmeter (see Fig. 6.3 and Table 6.1).
Note!
If transducers are replaced or added, the sensor module has to be
replaced or added, too (see section 6.10).
section
page
6.4.1 Transducers (ATEX Zone 1) - Direct Connection..............................................40
6.4.2 Transducers (ATEX Zone 1) - Connection via Junction Box .............................42
Connection of the Extension Cable with the Flowmeter ....................................42
Connection of the Extension Cable with the Junction Box ................................43
Connection of the Transducer Cable with the Junction Box..............................45
6.4.3 Transducers (ATEX Zone 2, FM Div. 2 and without Explosion Protection) -
Direct Connection..............................................................................................46
6.4.4 Transducers (ATEX Zone 2, FM Div. 2 and without Explosion Protection) -
Connection via Junction Box .............................................................................47
Connection of the Extension Cable with the Flowmeter ....................................47
Connection of the Extension Cable with the Junction Box ................................48
Connection of the Transducer Cable with the Junction Box..............................48
Attention!
For good high frequency shielding it is important to assure good con-
tact between the cable shield and the cap nut (and thus the housing).
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......