7 Installation of FLUXUS G709
7.6
Connection of the Inputs
Max. 4 temperature probes Pt100 (4-wire) can be connected to the flowmeter (option).
Configure the temperature inputs (see chapter 17).
7.6.1 Connection of a Current Input
An active current source has to be connected to a passive current input. A passive cur-
rent source can be connected directly to an active current input, or to a passive current
input via an external power supply.
Connection of a Passive Current Source
At full load (20 mA), a DC voltage of min. 13 V is available for the supply of a passive cur-
rent source.
Terminal TxA is on the upper, Txb on the lower terminal row (x = 1...4).
Fig. 7.9: Connection of an external passive current source
Table 7.10: Terminal assignment of the flowmeter
terminal
connection with
extension cable
connection without
extension cable
Txa
red
red
TxA
gray
red/blue
Txb
blue
white/blue
TxB
white
white
Sx
Pt100 shield
Pt100 shield
x = 1...4
Attention!
The terminals Txa and TxB must not be connected.
Attention!
Observe the polarity to avoid damage of the current source. The cur-
rent input could be destroyed by a permanent shorting of both termi-
nals.
+
-
Txb
TxA
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......