2 Safety Instructions according to Directive 94/9/EC
Type **K1N43, **G1N43, **H1N43, **M1N43, **P1N43
manufacturer
FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044;
II2G Ex q II T6…T3 T
a
-55…+140 °C
II2D Ex tD A21 IP65 TX
certification
IBExU04ATEX1011 X
type of protection
transducer:
powder filling
dust:
protection by enclosure tD
degree of protection
IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T3: -55…+140 °C
temperature class T4: -55…+125 °C
temperature class T5: -55…+90 °C
temperature class T6: -55…+75 °C
Type **M2E45, **P2E45
manufacturer
FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044;
II2G Ex q II T6…T2 T
a
-45…+225 °C
II2D Ex tD A21 IP65 TX
certification
IBExU04ATEX1011 X
type of protection
transducer:
powder filling
dust:
protection by enclosure tD
degree of protection
IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T2: -45…+225 °C
temperature class T3: -45…+190 °C
temperature class T4: -45…+125 °C
temperature class T5: -45…+90 °C
temperature class T6: -45…+75 °C
The service temperature at the cable entry must not
exceed the permissible operating temperature of 190 °C.
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......