62
7 Installation of FLUXUS G709
Fig. 7.3: Connection of extension cable and transducer cable (ATEX zone 1)
Connection of the Transducer Cable with the Junction Box
• Remove the blind plug from the junction box (see Fig. 7.2).
• Open the cable gland of the transducer cable. The compression part remains in the
cap nut.
• Push the transducer cable through cap nut, compression part and basic part.
• Prepare the transducer cable with the cable gland. Cut the outer shield and brush it
back.
• Insert the transducer cable in the junction box.
• Tighten the gasket ring side of the basic part in the junction box.
• Fix the cable gland by screwing the cap nut on the basic part.
• Connect the leads to the terminals of the junction box (see Fig. 7.2 and Table 7.3).
Table 7.3: Terminal assignment
terminal
connection
V
transducer
(core)
VS
transducer
(inner shield)
RS
transducer (inner shield)
R
transducer
(core)
shield terminal
equipotential
bonding terminal
TV TVS TG TG TRS TR
V
VS RS
R
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......