90
10 Basic Measurement
• Cut the damping mat to size, if necessary (see annex A, section Technical Data).
• Apply the a bead of ultrasound gel along the center line onto the inner side of the
damping mat.
• Spread the ultrasound gel evenly on the surface of the damping mat by means of a
Fig. 10.6: Application of ultrasound gel to the damping mat
• Position the damping mat on the pipe (see Fig. 10.3).
• Remove air pockets by means of the supplied roller.
• Mount the transducers onto the damping mat. Select the installation instruction corre-
sponding to the supplied transducer mounting fixture (see section 10.7).
• If the measuring point is close to a reflection point (e.g. flange), a damping mat has to
be installed there, if necessary (see Fig. 10.5).
• Repeat the steps for the installation of the damping mat.
Note!
When measuring in diagonal mode, a damping mat will be fixed addi-
tionally to the opposite side of the pipe (see Fig. 10.4).
Note!
Damping mats will be mounted to the complete circumference of the
pipe at the reflection point (see Fig. 10.5).
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......