128
13 Storing and Output of Measured Values
13.4.4 Storing of the Signal Amplitude
Select
SPECIAL FUNCTION\SYSTEM SETTINGS\STORING\STORE AMPLITUDE
.
13.4.5 Storing the Sound Velocity of the Medium
Select
SPECIAL FUNCTION\SYSTEM SETTINGS\STORING\STORE C-MEDIUM
.
13.4.6 Acoustic Signal
Per default, an acoustic signal will be emitted every time a measured value is stored or
transmitted to a PC or printer. The signal can be deactivated in
SPECIAL FUNC-
TION\SYSTEM SETTINGS\BEEP ON STORAGE
.
13.5 Available Data Memory
Max. 100 measuring data sets will be stored. The number of measuring data sets de-
pends on the total number of measured values stored in the precedent measuring data
sets.
If the data memory is empty and a measurement is started with one physical quantity on
one measuring channel without storing of the totalizer, approx. 100 000 measured values
can be stored. The available data memory can be displayed:
If
ON
is selected and the data memory is activated, the
amplitude of the measured signal will be stored to-
gether with the measured values. Press ENTER.
If
ON
is selected and the data memory is activated, the
sound velocity of the medium will be stored together
with the measured values. Press ENTER.
Select
OFF
to deactivate the acoustic signal,
ON
to acti-
vate it. Press ENTER.
The time on which the memory will be full can be dis-
played during the measurement.
Scroll through the displays of the upper line with key
during measurement.
Select
SPECIAL FUNCTION\INSTRUM. INFORM.
Press ENTER.
Type and serial number of the flowmeter will be dis-
played in the upper line.
The available data memory will be displayed in the low-
er line (here: 18 327 measured values can be stored).
Press twice key BRK to return to the main menu.
Store Amplitude
off
>ON<
Store c-Medium
off
>ON<
Beep on storage
>ON<
off
FULL= 26.01/07:39
54.5
m3/h
DISP
Special Function
Instrum. Inform.
G70X-XXXXXXXX
Free:
18327
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......