104
11 Displaying the Measured Values
11.5 Transducer Distance
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses in the
upper line, followed by the entered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during measurement (e.g. due to tem-
perature fluctuations).
A deviation from the optimum transducer distance (here: -0.4 mm) will be compensated
internally.
R
flow profile
information about the flow profile based on the Reynold's number
T
fully turbulent flow profile
L
fully laminar flow profile
the flow is in the transition range between laminar and turbulent flow
?
unknown, can not be calculated
F
flow velocity
comparison of the measured flow velocity with the flow limits of the system
ok, flow velocity is not in a critical range
the flow velocity is higher than the actual limit
the flow velocity is lower than the actual cut-off flow (even if it is not set to zero)
0
the flow velocity is in the offset range of the measuring method
?
unknown, can not be measured
By pressing key
, it is possible to scroll to the display of
the transducer distance during measurement.
Note!
Never change the transducer distance during measurement!
L=(51.2) 50.8 mm
54.5
m3/h
DISP
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......