157
18 Outputs
18.2 Error Value Delay
The error value delay is the time interval after which the error value will be transmitted to
the output in case no valid measured values are available. The error value delay can be
entered in the program branch
OUTPUT OPTIONS
if this menu item has been previously
activated in the program branch
SPECIAL FUNCTION
. If the error value delay is not en-
tered, the damping factor will be used.
18.3 Activation of an Analog Output
18.3.1 Measuring Range of the Analog Outputs
After an analog output has been activated in the program branch
OUTPUT OPTIONS
, the
measuring range of the source item has to be entered.
Select
SPECIAL FUNCTION\SYSTEM SETTINGS\DIA-
LOGS/MENUS\ERROR-VAL. DELAY
Select
DAMPING
if the damping factor is to be used as er-
ror value delay. Select
EDIT
to activate the error value de-
lay request.
From now on, the error value delay can be entered in the
program branch
OUTPUT OPTIONS
.
Note!
An output can only be activated in the program branch
OUTPUT OP-
TIONS
if it has been previously installed.
Select in the program branch
OUTPUT OPTIONS
the chan-
nel for which an output has to be activated. Press ENTER.
This display will not be indicated, if the flowmeter has only
one measuring channel.
Press ENTER until
CURRENT LOOP
is displayed. Select
YES
to activate the output. Press ENTER.
Select
SIGN
if the sign of the measured values has to be
considered for the output.
Select
ABSOLUT
if the sign does not need to be consid-
ered.
Enter in
ZERO-SCALE VAL.
the lowest measured value
expected. The unit of measurement of the source item will
be displayed.
ZERO-SCALE VAL.
is the measured value corresponding
to the lower limit of the output range as defined in section
18.1.1.
Error-val.delay
>DAMPING<
edit
Error-val. delay
10
s
Output Options
for Channel
A:
Current Loop
I1: no
>YES<
Meas.Values
>ABSOLUT<
sign
Zero-Scale Val.
0.00
m3/h
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......